1. <source id="vki21"></source>
        2. OEM Icon

          OEM 无创血压技术

          我们的技术。你的解决方案。

          功能与Adv A+相当,但体积更小,功耗更低,所有硬件集成在一起,是急救医疗服务和手持医疗设备的理想选择。 1) 体积小 2) RMT和TMT技术可选 3) 符合医疗设备第三版和Rohs要求 4) RS232和TTL通讯协议可选 5) 易安装 6) Adv A+模块的紧凑型

          Adv A+ CS 模块

          技术规格

          Physical Dimensions

          SPECIFICATIONS VALUES
          Width 1.9 in [48.5 mm]
          Length 2.56 in [65.0 mm]
          Height 0.89 in [22.5 mm]
          Weight 0.1875 lbs [85 g]

          Measurement

          SPECIFICATIONS VALUES
          Technique Oscillometric
          NIBP Accuracy Meets ANSI/AAMI SP10, EN 1060-4, ISO 81060-2
          Patient Applications Adult, Pediatric, Neonatal
          Systolic Range
          • Adult: 40-260 mmHg
          • Pediatric: 40-160 mmHg
          • Neonatal: 40-130 mmHg
          Diastolic Range
          • Adult: 20-200 mmHg
          • Pediatric: 20-120 mmHg
          • Neonatal: 20-100 mmHg
          Pulse Rate Range 30-220 BPM (Beats per min)
          Pulse Rate Accuracy ±2% or ±3 BPM, whichever is greater
          Transducer Accuracy ± 3 mmHg over full range in operating conditions
          Operating Modes Manual
          Operating Conditions 0?C to 50?C, 15% to 95%, non-condensing humidity
          Storage Conditions -20?C to +65?C, up to 90% RH non-condensing

          Electrical

          SPECIFICATIONS VALUES
          Operating Voltage 5.3-14.5 VDC (High), 1.8-9.0 VDC (Low)
          Power Consumption Using 6 VDC Nominal (High) or 3.3 VDC Nominal (Low)
          • Sleep: 2.0 mW (High), 3.0 mW (Low)
          • Idle: 348 mW (High), 430 mW (Low)
          • Measurement: 1.3 W (High), 1.4 W (Low)
          • Inflation: 3.7 W (High), 3.4 W (Low)
          Communication Protocols Serial RS232 or TTL (3.3V or 5V)
          Safety & Regulatory Standards IEC 60601-1, IEC/EN60601-2-30, AAMI SP10, EN1060-1, EN1060-3, EN1060-4, IEC 80601-2-30, ISO 81060-2

          Advantage Technologies

          SPECIFICATIONS VALUES
          SMT (Standard Motion Tolerance) Yes
          RMT-S (Standard R-wave Motion Tolerance) Yes
          TMT-S (Standard Transport Motion Tolerance) Yes
          VET (Veterinary Monitoring) Yes
          HDM (Hemodialysis Monitoring) Yes

          Customer Support

          SPECIFICATIONS VALUES
          Standard Support Yes
          Custom Support 10 hours*
          Custom Design Additional Charge

          Warranty

          SPECIFICATIONS VALUES
          Time 1 year
          Cycles 30000